Zero point adjustment

All measuring devices are calibrated to state-of-the-art technology. The zero point determined in this way is 
imprinted on the nameplate.
Calibration takes place under reference conditions. 

Page 12 ff. 

Therefore, a zero point adjustment is generally 

not

 required!

Experience shows that the zero point adjustment is advisable only in special cases:
• When the highest measuring accuracy is required and the flow rates are very low.
• Under extreme process or operating conditions (e.g. very high process temperatures or very high viscosity 

fluids).

Please note the following before carrying out the adjustment:
• The adjustment can only be performed with fluids that have no gas or solid contents.
• Zero point adjustment is performed with the measuring tubes completely filled and at zero flow (v = 0 m/s). 

This can be achieved, for example, with shutoff valves upstream and/or downstream of the sensor or by 
using existing valves and gates.
– Normal operation 

 valves 1 and 2 open

– Zero point adjustment 

with

 pump pressure 

 valve 1 open / valve 2 closed

– Zero point adjustment 

without

 pump pressure 

 valve 1 closed / valve 2 open

a0003601

Zero point adjustment and shutoff valves

Inlet and outlet runs

There are no installation requirements regarding inlet and outlet runs.

Length of connecting cable

Max. 20 meters (remote version)

System pressure

It is important to ensure that cavitation does not occur, because it would influence the oscillation of the 
measuring tube. No special measures need to be taken for fluids which have properties similar to water under 
normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is 
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to 
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects 
can be prevented when system pressure is sufficiently high.

Therefore, the following locations should be preferred for installation:
• Downstream from pumps (no danger of vacuum)
• At the lowest point in a vertical pipe

1

2

Содержание Proline Promass 80H

Страница 1: ...elevant safety aspects Secondary containment up to 25 bar Pressure Equipment Directive SIL 2 Your benefits The Promass measuring devices make it possible to simultaneously record several process varia...

Страница 2: ...nstallation instructions 15 Inlet and outlet runs 18 Length of connecting cable 18 System pressure 18 Operating conditions Environment 19 Ambient temperature range 19 Storage temperature 19 Degree of...

Страница 3: ...t the inlet and outlet For the Promass H the system balance is created by a counterweight that runs parallel to the measuring tube This counterweight oscillates in antiphase to the measuring tubes and...

Страница 4: ...peration with Touch control Application specific Quick Setup Mass flow volume flow density and temperature measurement as well as calculated variables e g fluid concentrations H a0003677 Single bent t...

Страница 5: ...for fluid temperatures up to 350 C Nominal diameters DN 25 50 80 Tube material Alloy C 22 M a0003676 Robust sensor for extreme process pressures high requirements for the secondary containment and fl...

Страница 6: ...ally isolated time constant selectable 0 05 to 100 s full scale value selectable temperature coefficient typically 0 005 o f s C resolution 0 5 A Active 0 4 to 20 mA RL 700 for HART RL 250 Passive 4 t...

Страница 7: ...g mode Totalizer control Bus address can be configured via miniature switches or via the local display optional Available output combination Page 10 PROFIBUS PA interface PROFIBUS PA in accordance wit...

Страница 8: ...n alarm Current output Failsafe mode selectable e g in accordance with NAMUR Recommendation NE 43 Pulse frequency output Failsafe mode selectable Status output Promass 80 Nonconductive in the event of...

Страница 9: ...verse polarity protection Terminal No 27 DP B PA FF MODBUS RS485 B PA FF with reverse polarity protection e Ground terminal for signal cable shield fieldbus cable RS485 line f Service adapter for conn...

Страница 10: ...Terminal No inputs outputs Order version 20 21 22 23 24 25 26 27 Fixed communication boards permanent assignment 83 A Frequency output Current output HART 83 B Relay output Relay output Frequency out...

Страница 11: ...cy output Status input MODBUS RS485 83 P Current output Frequency output Status input PROFIBUS DP 83 V Relay output 2 Relay output 1 Status input PROFIBUS DP 83 W Relay output Current output 3 Current...

Страница 12: ...ails HistoROM S DAT exchangeable data storage chip with sensor specific data nominal diameter serial number calibration factor zero point etc Promass 83 Lasting min 1 power cycle EEPROM and T DAT save...

Страница 13: ...r Max measured error 0 125 1 8 kg h 8000 kg h 100 0 15 Density liquid 1 g cc 1 kg l Standard calibration 0 02 g cc Special density calibration optional calibration range 0 8 to 1 8 g cc 5 to 80 C 0 0...

Страница 14: ...lue x 100 o r Repeatability 0 05 1 2 1 8 kg h 8000 kg h 100 0 061 Density measurement liquid 1 g cc 1 kg l 0 0005 g cc Temperature measurement 0 25 C 0 0025 T T fluid temperature in C Influence of flu...

Страница 15: ...vy sensors Mounting location Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors Therefore avoid the following mounting locations in the pipe installation...

Страница 16: ...buildup Horizontal The transmitter can be installed in any orientation in a horizontal pipe run m To ensure that the maximum permitted ambient temperature for the transmitter 20 to 60 C optionally 40...

Страница 17: ...the sensor and transmitter and the connection housing of the remote version always remain free of insulating material Note that a certain orientation might be required depending on the fluid temperat...

Страница 18: ...ormal operation valves 1 and 2 open Zero point adjustment with pump pressure valve 1 open valve 2 closed Zero point adjustment without pump pressure valve 1 closed valve 2 open a0003601 Zero point adj...

Страница 19: ...Cl 150 Cl 300 JIS 10K 20K Pressure ranges of secondary containment DN 8 to 15 25 bar DN 25 to 50 16 bar Warning In case a danger of measuring tube failure exists due to process characteristics e g wit...

Страница 20: ...id density kg m3 d inside diameter of measuring tubes m K to K3 constants depending on nominal diameter To compute the pressure loss for gases always use the formula for Re 2300 DN d m K K1 K3 8 8 51...

Страница 21: ...connection housing remote version II2G Zone 1 Page 23 Transmitter wall mount housing non Ex zone and II3G Zone 2 Page 24 Flange connections EN DIN ASME B16 5 JIS Page 25 Promass purge connections seco...

Страница 22: ...r compact version stainless steel field housing a0002245 Dimensions Remote version a0002516 A B C 220 153 171 All dimensions in mm A B C Esc E B C A DN A B C 8 118 5 137 5 127 15 118 5 137 5 127 25 11...

Страница 23: ...ons in mm Esc E Nicht unter Spannung ffnen K e e p c o v er t i g h t w h i l e c i r c u i t s a r e a l i v e N e p a s o u v r i r l a p p a r e i l s o u s t e n s i o n K e e p c o v e r t i g h...

Страница 24: ...ser Transmitter wall mount housing non Ex zone and II3G Zone 2 a0001150 A B C D E F G H J 215 250 90 5 159 5 135 90 45 50 81 K L M N O P Q R 53 95 53 102 81 5 11 5 192 8xM5 All dimensions in mm Esc E...

Страница 25: ...ghness flange Ra 3 2 to 6 3 m DN G L N S LK U di 8 1 88 9 336 4 x 15 7 12 8 60 5 15 70 8 51 15 88 9 440 4 x 15 7 12 8 60 5 15 70 12 00 25 108 0 580 4 x 15 7 15 1 79 2 26 70 17 60 40 127 0 794 4 x 15 7...

Страница 26: ...langes Weight information in kg Flange JIS B2238 20K 1 4301 304 fluid wetted parts zirconium 702 Surface roughness flange Ra 3 2 to 6 3 m DN G L N S LK U di 8 1 95 336 4 x 15 14 70 15 00 8 51 15 95 44...

Страница 27: ...ocess connections Flanges EN 1092 1 DIN 2501 ASME B16 5 JIS B2238 stainless steel 1 4301 304 fluid wetted parts zirconium 702 Measuring tubes Zirconium 702 R 60702 Seals Welded process connections wit...

Страница 28: ...k Setup menus for straightforward commissioning Promass 83 Local operation with three optical keys S O F Application specific Quick Setup menus for straightforward commissioning Language groups Note T...

Страница 29: ...l the requirements of the following specifications Certified in accordance with PROFIBUS Profile Version 3 0 device certification number available on request The device can also be operated with certi...

Страница 30: ...ring information and information on the order codes on request Accessories Various accessories which can be ordered separately from Endress Hauser are available for the transmitter and the sensor Prom...

Страница 31: ...6 en Description of Device Functions Promass 83 BA060D 06 en Description of Device Functions Promass 83 FOUNDATION Fieldbus BA066D 06 en Description of Device Functions Promass 83 PROFIBUS DP PA BA064...

Страница 32: ...Endress Hauser GmbH Co KG Instruments International Colmarer Str 6 79576 Weil am Rhein Deutschland Tel 49 76 21 9 75 02 Fax 49 76 21 9 75 34 5 www endress com info ii endress com TI074D 06 en 12 06 7...

Отзывы: