Operation options
Proline Promag W 400
50
Hauser
Switching on the keypad lock
The keypad lock is switched on automatically:
• Each time the device is restarted.
• If the device has not been operated for longer than one minute in the measured value
display.
1. The device is in the measured value display.
Press the key for longer than 2 seconds.
A context menu appears.
2. In the context menu, select the
Keylock on
option.
The keypad lock is switched on.
Switching off the keypad lock
1. The keypad lock is switched on.
Press the key for longer than 2 seconds.
A context menu appears.
2. In the context menu, select the
Keylock off
option.
The keypad lock is switched off.
8.3
Access to the operating menu via the Web browser
8.3.1
Function range
Thanks to the integrated Web server the device can be operated and configured via a Web
browser. The operating menu structure is the same as in the local display.
8.3.2
Prerequisites
Hardware
Connecting cable
Standard Ethernet cable with RJ45 connector
Computer
RJ45 interface
Measuring device:
Web server must be enabled; factory setting: ON
IP address
If the IP address of the device is not known:
• The IP address can be read out via local operation:
"Diagnostics" menu → Device info → IP address
• Communication with the Web server can be established via the standard IP
address 192.168.1.212.
The DHCP function is enabled in the device at the factory, i.e. the device
expects an IP address to be assigned by the network. This function can be
disabled and the device can be set to the standard IP address 192.168.1.212:
set top DIP switch No. 2 from OFF → ON.