Electrical connection
Proline Promag W 400
28
Hauser
Operation in zones of severe electrical interference
Grounding is by means of the ground terminal provided for the purpose inside the connection
housing. The stripped and twisted lengths of cable shield to the ground terminal must be as
short as possible.
Cable diameter
• Cable glands supplied:
– For standard cable: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
– For reinforced cable: M20 × 1.5 with cable 9.5 to 16 mm (0.37 to 0.63 in)
• (Plug-in) spring terminals for wire cross-sections 0.5 to 2.5 mm
2
(20 to 14 AWG)
7.1.3
Terminal assignment
Transmitter
Supply voltage
Order code for "Power supply"
Terminal numbers
1 (L+/L)
2 (L-/N)
Option
L
(wide range power unit)
AC100 to 240 V
AC/DC24 V
Signal transmission 0-20 mA/4-20 mA HART with additional outputs and inputs
Order code for
"Output" and "Input"
Terminal numbers
Output 1
Output 2
Output 3
Input
26 (+)
27 (-)
24 (+)
25 (-)
22 (+)
23 (-)
20 (+)
21 (-)
Option
H
•
4-20 mA HART
(active)
•
0-20 mA
(active)
Pulse/frequency
output
(passive)
Switch output
(passive)
-
Option
I
•
4-20 mA HART
(active)
•
0-20 mA
(active)
Pulse/frequency/
switch output
(passive)
Pulse/frequency/
switch output
(passive)
Status input
Option
J
•
4-20 mA HART
(active)
•
0-20 mA
(active)
Permanently assigned
Pulse output adjusted
(passive)
Pulse/frequency/
switch output
(passive)
Status input