Oxymax W COS61
Installation
Hauser
9
a0004102
Fig. 4:
Measuring system (example)
1
Immersion assembly CYA611
2
Junction box VS (optional)
3
Transmitter Liquisys M COM253
4
Oxygen sensor COS61
3.3.2
Installing a measuring point
!
Note!
For immersed operation, install the individual modules away from the basin on a solid base. Only
carry out the final installation at the intended installation location.
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning
function).
3.
Install and connect the oxygen sensor.
4.
Install an immersion or a suspension assembly (if used) into the process.
"
Caution!
• For immersed operation, the sensor must be installed in an immersion assembly (e.g. CYA611).
Do not install the sensor suspended from the cable.
• Screw the sensor into the assembly so that the cable is not twisted.
• Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
• Select the installation location so that there is easy access for later calibration.
#
Warning!
When using metallic assemblies and installation equipment, comply with national grounding
regulations.