Troubleshooting
Oxymax W COS61
26
Hauser
3.
Immerse the sensor into the water and wait a sufficient period of time (15 min. for oxygen
depletion).
The display drops to around 0 mg/l (0 %SAT).
Depending on the conditions (contact surface water/air), the zero solution is stable for up to 12
hours.
!
Note!
If there are deviations from the reference values, follow the troubleshooting instructions or contact
your sales office.
9.3
Spare parts
9.4
Return
If the device requires repair, please send it
cleaned
to the sales center responsible.
Please use the original packaging, if possible.
Please enclose the completed "Declaration of contamination" (copy the second last page of these
Operating Instructions) with the packaging and the transportation documents.
No repair without
completed "Declaration of contamination"!
9.5
Disposal
The device contains electronic components and must therefore be disposed of in accordance with
regulations on the disposal of electronic waste.
Please observe local regulations.
Position
Spare parts kit
order no.
a0004166
Fig. 18:
1
Sensor
see product
structure
2
Sealing ring
– 2 pieces
51518597
3
Sensor cap (fluorescence cap)
51518598
4
Protection guard
on request
without fig.
Zero solution
– 3 units to produce 3 x 1 liter oxygen free solution
50001041