Wiring
Oxymax W COS61
14
Hauser
4
Wiring
#
Warning!
• The electrical connection must only be carried out by an electrical technician.
• The electrical technician must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable
before beginning
the connection work.
4.1
Direct connection to the transmitter
• COS61 works with the additional 15 V power supply from Liquisys.
• Check the wiring or change the wiring as follows (see Operating Instructions COM223/253):
– Connect the yellow cable cores to terminal 85 (+15 V).
– Connect terminals 86 and 0 with an
additional
gray wire.
4.1.1
Field installation
You connect the sensor directly to the transmitter (COM253-WX/WS) by means of the special
measuring cable with the SXP connector (
4.1.2
Panel mounting
• Remove the SXP connector (transmitter side!) from the cable.
• Refer to the following table for the cable assignment and the assigned terminals for Liquisys M
COM223-WX/WS.
• Please note that the cable assignment changes depending on the sensor version (fixed cable or
TOP68 connection).
a0004118
Fig. 14:
SXP connector
Terminal
COM223
Sensor with fixed cable (OMK)
Sensor with TOP68 connection (CYK71)
Core
Assignment
Core
Assignment
85
YE
+U
B
YE
+U
B
0, 86
GY
0 V
WH
0 V
96
PK
Comm. (digital)
GN
Communication (digital)
97
BU
Comm. (digital)
BN
Communication (digital)
88
BN
–U
B
Coax,
inner
–U
B
Additional connection between terminals 86 and 0 of the transmitter necessary.