Device description
Oxymax W COS51D
18
Hauser
5.3.3
Calibration in air
1.
Remove the sensor from the medium.
2.
Clean the outside of the sensor with a damp cloth.
Then dry the sensor membrane e.g. by using a tissue.
3.
Then wait while the sensor adjusts to the temperature of the ambient air. This takes about 20
minutes. Check that the sensor is not in direct sunlight during this time.
4.
If the measured value display on the transmitter is stable, carry out the calibration in
accordance with the Operating Instructions of the transmitter.
5.
Place the sensor in the medium again.
!
Note!
Make sure you comply with the instructions for calibration in the Operating Instructions of the
transmitter.
5.3.4
Calculation example for the calibration value
As a check, you can calculate the expected calibration value (transmitter display) as shown in the
following example (salinity is 0).
1.
Determine:
– The ambient temperature for the sensor (air temperature for "air" calibration method,water
temperature for "air-saturated water" calibration type)
– the altitude above sea level
– the current air pressure
L
(=rel. air pressure to sea level) at the time of calibration. (If
undeterminable, use 1013 hPa (407 inH
2
O) for an approximate calculation.)
2.
Define:
– the saturation value
S
acc. to the first table
– the factor
K
acc. to the second table
° C / °F
S
[mg/l=ppm]
° C / °F
S
[mg/l=ppm]
° C / °F
S
[mg/l=ppm]
° C / °F
S
[mg/l=ppm]
0 / 32
14.64
11 / 52
10.99
21 / 70
8.90
31 / 88
7.42
1 / 34
14.23
12 / 54
10.75
22 / 72
8.73
32 / 90
7.30
2 / 36
13.83
13 / 55
10.51
23 / 73
8.57
33 / 91
7.18
3 / 37
13.45
14 / 57
10.28
24 / 75
8.41
34 / 93
7.06
4 / 39
13.09
15 / 59
10.06
25 / 77
8.25
35 / 95
6.94
5 / 41
12.75
16 / 61
9.85
26 / 79
8.11
36 / 97
6.83
6 / 43
12.42
17 / 63
9.64
27 / 81
7.96
37 / 99
6.72
7 / 45
12.11
18 / 64
9.45
28 / 82
7.82
38 / 100
6.61
8 / 46
11.81
19 / 66
9.26
29 / 84
7.69
39 / 102
6.51
9 / 48
11.53
20 / 68
9.08
30 / 86
7.55
40 / 104
6.41
10 / 50
11.25
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
0
1.000
550 / 1800
0.938
1050 / 3450
0.885
1550 / 5090
0.834
50 / 160
0.994
600 / 1980
0.932
1100 / 3610
0.879
1600 / 5250
0.830
100 / 330
0.988
650 / 2130
0.927
1150 / 3770
0.874
1650 / 5410
0.825
150 / 490
0.982
700 / 2300
0.922
1200 / 3940
0.869
1700 / 5580
0.820
200 / 660
0.977
750 / 2460
0.916
1250 / 4100
0.864
1750 / 5740
0.815
Содержание Oxymax W COS51D
Страница 1: ...BA413C 07 en 10 08 71083533 Operating Instructions Oxymax W COS51D Digital sensor for dissolved oxygen...
Страница 31: ...Oxymax W COS51D EC Declaration of Conformity Endress Hauser 31 11 EC Declaration of Conformity...
Страница 34: ...Oxymax W COS51D 34 Endress Hauser...
Страница 36: ...www endress com worldwide BA413C 07 en 10 08 Printed in Germany FM SGML 6 0 DT...