Oxymax W COS51D
Installation
Hauser
9
3.3.2
Installing a measuring point
!
Note!
For immersed operation, install the individual modules away from the basin on a solid base. Only
carry out the final installation at the intended installation location.
For a complete installation of a measuring point, proceed as follows:
1.
Install a retractable or a flow assembly (if used) into the process.
2.
Connect the water supply to the rinse connections (if you use an assembly with cleaning
function).
3.
Install and connect the oxygen sensor.
4.
Install an immersion or an suspension assembly (if used) into the process.
"
Caution!
• For immersed operation, the sensor must be installed in an immersion assembly (e.g. CYA611).
Do not install the sensor suspended from the cable.
• Screw the sensor into the assembly so that the cable is not twisted.
• Avoid exerting excessive tensile force on the cable (e.g. from jerky pulling).
• Select the installation location so that there is easy access for later calibration.
#
Warning!
When using metallic assemblies and installation equipment, comply with national grounding
regulations.
a0006735
Fig. 3:
Measuring system (example)
1
Digital oxygen sensor Oxymax W COS51D
2
Retractable assembly COA451
3
Liquiline M CM42
4
Measuring cable CYK10
Содержание Oxymax W COS51D
Страница 1: ...BA413C 07 en 10 08 71083533 Operating Instructions Oxymax W COS51D Digital sensor for dissolved oxygen...
Страница 31: ...Oxymax W COS51D EC Declaration of Conformity Endress Hauser 31 11 EC Declaration of Conformity...
Страница 34: ...Oxymax W COS51D 34 Endress Hauser...
Страница 36: ...www endress com worldwide BA413C 07 en 10 08 Printed in Germany FM SGML 6 0 DT...