Oxymax W COS51D
Technical data
Hauser
29
10
Technical data
10.1
Input
Measured variable
Dissolved oxygen [mg/l,
μ
g/l, ppm, ppb, % SAT or hPa]
Measuring range
0.01 to 100 mg/l
0.00 to 1000 % SAT
0 to 2000 hPa
10.2
Performance characteristics
Response time
• COS51D-***0* (black membrane cap for normal response time):
– t
90
: 3 minutes
– t
98
: 8minutes (each at 20 °C / 68 °F)
• COS51D-***1* (white membrane cap for fast response time):
– t
90
: 0.5 minutes
– t
98
: 1.5 minutes (each at 20 °C / 68 °F)
Reference operating
conditions
Signal current in air
1)
• COS51D-***0* (black membrane cap):
approx. 300 nA
• COS51D-***1* (white membrane cap):
approx. 1100 nA
Polarization time
< 60 minutes
Longtime drift
Zero current
< 0.1 % of the current in air
Measured value resolution
0.01 mg/l (0.01 ppm)
Maximum measured error
±1 % of measured value
2)
Oxygen intrinsic
consumption
• COS51D-***0*:
Approx. 90 ng/h in air at 25 °C (77 °F)
• COS51D-***1*:
Approx. 270 ng/h in air at 25 °C (77 °F)
Repeatability
±1 % of measured value
Reference temperature:
25 °C (77 °F)
Reference pressure:
1013 hPa (15 psi)
1)
at reference operating conditions
Zero-point drift:
< 0.1 % per week at 30 °C (86 °F)
Measuring range drift:
< 0.1 % per week at 30 °C (86 °F)
1)
1)
under constant conditions each
2)
In accordance with IEC 746-1 at nominal operating conditions
Содержание Oxymax W COS51D
Страница 1: ...BA413C 07 en 10 08 71083533 Operating Instructions Oxymax W COS51D Digital sensor for dissolved oxygen...
Страница 31: ...Oxymax W COS51D EC Declaration of Conformity Endress Hauser 31 11 EC Declaration of Conformity...
Страница 34: ...Oxymax W COS51D 34 Endress Hauser...
Страница 36: ...www endress com worldwide BA413C 07 en 10 08 Printed in Germany FM SGML 6 0 DT...