Oxymax W COS51D
Device description
Hauser
19
3.
Calculate the calibration value
C
:
C = S
.
K
.
L
Example
• Air calibration at 18°C (64 °F), altitude 500 m (1650 ft) above sea level, air pressure 1009 hPa
(405 inH
2
O) =1.009 bar
• S = 9.45 mg/l, K = 0.943, L = 1.009
Calibration value C = 9.17 mg/l.
!
Note!
You do not need factor K from the table if your device returns the absolute air pressure L
abs
(location-dependent air pressure) as the measured value.
Thus, the formula for calculation is: C = S
.
L
abs
.
250 / 820
0.971
800 / 2620
0.911
1300 / 4270
0.859
1800 / 5910
0.810
300 / 980
0.966
850 / 2790
0.905
1350 / 4430
0.854
1850 / 6070
0.805
350 / 1150
0.960
900 / 2950
0.900
1400 / 4600
0.849
1900 / 6230
0.801
400 / 1320
0.954
950 / 3120
0.895
1450 / 4760
0.844
1950 / 6400
0.796
450 / 1480
0.949
1000 / 3300
0.890
1500 / 4920
0.839
2000 / 6560
0.792
500 / 1650
0.943
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K
Altitude
[m / ft]
K