Oxymax W COS51D
Device description
Hauser
15
5
Device description
5.1
Sensor design
The sensor consists of the following function units:
• Sensor shaft
• Protection basket
!
Note!
• Alternatively to the protection basket, you can use a spray head COR 3 (optional, see
"Accessories") for use in immersed operation with cleaning function.
• The membrane cap screwed onto the sensor head is filled with electrolyte. The screw connection
seals it from the medium.
• The membrane which is in contact with the medium is pretensioned in the factory.
a0006935
Fig. 15:
Sensor design
1
Memosens Plug-in head
2
Threaded connection NPT ¾"
3
Threaded connection G1
4
Sensor shaft
5
Protection basket
6
Gold cathode
7
Membrane
8
Electrolyte
9
Anode
a0004149
Fig. 16:
Sensor head, cutaway drawing
7
Membrane
8
Electrolyte
9
Anode
10
Reference electrode
11
Threaded connection for protection basket
12
Sealing ring
13
Threaded connection for membrane cap
14
Membrane cap
a0004150
Fig. 17:
Sensor head, top view
6
Cathode
1
3
4
5
6
7
8
9
2
10
11
12
13
14
7
8
9
10
6
Содержание Oxymax W COS51D
Страница 1: ...BA413C 07 en 10 08 71083533 Operating Instructions Oxymax W COS51D Digital sensor for dissolved oxygen...
Страница 31: ...Oxymax W COS51D EC Declaration of Conformity Endress Hauser 31 11 EC Declaration of Conformity...
Страница 34: ...Oxymax W COS51D 34 Endress Hauser...
Страница 36: ...www endress com worldwide BA413C 07 en 10 08 Printed in Germany FM SGML 6 0 DT...