Liquiline CM44x Modbus Guideline V1.0
Hauser
20
© Copyright Dez. 2011
6.2.
Diagnostic messages
The device features via Modbus register detailled informations to current diagnostic messages with the highest
priority and for the last fallen diagnostic message.
The diagnostic codes and the related remedy measures are described in the operation manual BA477C.
6.2.1.
Device related diagnostic
Current active diagnostic message with the highes priority:
Parameter
Input
Register
Datatyp
Information
DEVICE_CURRENT_DIAG_MODUL
504
UNSIGNED16 Slot number of the module.
DEVICE_CURRENT_DIAG_PORT
505
UNSIGNED8
Connector on the module
DEVICE_CURRENT_DIAG_NE107_STATUS
506
UNSIGNED8
Bit 0: Failure (F)
Bit 1: Function Check (F)
Bit 2: Maintenance required(M)
Bit 3: Out of specification (S)
Bit 4-7: reserviert = 0
DEVICE_CURRENT_DIAG_CODE
507
UNSIGNED16 Diagnostic code (0..999)
DEVICE_CURRENT_DIAG_TIMESTAMP
508
UNSIGNED32 Timestamp (seconds since 1.1.1970)
DEVICE_NUMBER_ADDITIONAL_DIAG
510
UNSIGNED8
Number of further active diagnostic
messages with lower priority. A full list
of all messages is shown on the local
display.
Last fallen diagnostic message (the reason for the message is not given any more, therefor the diagnostic message is
withdrawn)
Parameter
Input
Register
Datentyp
Info
DEVICE_LAST_DIAG_MODUL
512
UNSIGNED8
See above
DEVICE_LAST_DIAG_PORT
513
UNSIGNED8
DEVICE_LAST_DIAG_NE107_STATUS
514
UNSIGNED8
DEVICE_LAST_DIAG_CODE
515
UNSIGNED16
DEVICE_LAST_DIAG_TIMESTAMP
516
UNSIGNED32