Liquiline CM44x Modbus Guideline V1.0
Hauser
14
© Copyright Dez. 2011
5.
Commissioning
5.1.
Local configuration of the device
All settings can be done at the display of the device or remotely via the service interface.
For the service interface connect the Commubox FXA291 and use the software FieldCare.
Note:
For further information please refer to the following documentation:
BA473C: Liquistation CSF22 Operating instructions (Commissioning).
BA476C: Liquistation CSF22 Operating instructions (Operation & settings).
5.2.
Configuration of the Modbus RS485 interface
5.2.1.
Settings
Settings
Default settings
selection
Bus address
247
1 ... 247
Transmission mode RTU
RTU, ASCII
Baud rate/bps
19200
1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Parity
even (1 Stopbit)
even (1 Stopbit), Uneven (1 Stopbit), non (2 Stopbits)
Path: Menu/Setup/General settings/Extended setup /Modbus/Settings:
5.2.2.
Selecting Bus address and Identnumber
Liquistation CSF22 supports two options to set the Modbus Address:
1. by Menu
Software-Addressing
2. by Modbus
Software-Addressing
Step 1
Step 2
Step 3
Step 4