The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-section
Maximum cable length
[mm
2
]
[AWG]
[m]
[ft]
0.5
20
70
230
0.75
18
100
328
1.0
17
100
328
1.5
16
200
656
2.5
14
300
984
Performance characteristics
reference operating
conditions
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the
Applicator
Maximum measured error
Base accuracy
Mass flow
±0.2 % o.r.
Volume flow
±0.3 % o.r.
Density
±20 kg/m³ (±0.02 SGU)
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Zero point stability
DN
Zero point stability
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0.20
0.007
15
½
0.65
0.024
25
1
1.80
0.066
40
1½
4.50
0.165
50
2
7.0
0.257
Flow values
Flow values as turndown parameter depending on nominal diameter.