LPGmass
30
Hauser
Efficient diagnostics increase measurement availability
• Troubleshooting measures can be called up via the operating tools
• Diverse simulation options
• Status indicated by several light emitting diodes (LEDs) on the electronic module in the housing
compartment
Service interface
Via service interface (CDI)
This communication interface is present in the following device version:
Order code for "Output", option
M
: Modbus RS485
Modbus RS485
1
2
3
A0030216
1
Service interface (CDI) of measuring device
2
Commubox FXA291
3
Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
Certificates and approvals
CE mark
The measuring system is in conformity with the statutory requirements of the applicable EU
Directives. These are listed in the corresponding EU Declaration of Conformity along with the
standards applied.
Hauser confirms successful testing of the device by affixing to it the CE mark.
C-Tick symbol
The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
Ex approval
The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Hauser sales center.
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex ia
Category (ATEX)
Type of protection
II2G
Ex ia IIC T6…T1 Gb
II1/2G, II2D
Ex ia IIC T6…T1 Ga/Gb
Ex tb IIIC Txx °C Db
II2G, II2D
Ex ia IIC T6…T1 Gb
Ex tb IIIC Txx °C Db