Liquiphant M/S with electronic insert FEL58 + Nivotester FTL325N
24
Hauser
Density
Set the density at the right switch:
Liquid density Function Switch position
Comment
>0.7 kg/dm³
>0.7
Top
(See
A
in the graphic below)
Standard setting;
Always use if possible
>0.5 kg/dm³
>0.5
Bottom
(See
B
in the graphic below)
Special settings;
Extremely light liquids (e.g.: liquefied natural gas)
U= 0 V
B
A
ρ
ρ
1
1
2
2
A0026156
A
Standard setting (density >0.7 kg/dm³)
A1 1 l (0.264 gal) or 1 dm³ (61.02 in³)
A2 >0.7 kg (1.54 lbs)
B
Special setting (density >0.5 kg/dm³)
B1 1 l (0.264 gal) or 1 dm³ (61.02 in³)
B2 >0.5 to 0.7 kg (1.10 to 1.54 lbs)