Diagnostics and troubleshooting
Liquiline CM44P
158
Hauser
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
950
Process
temperature
F
On
On
C
Concentration table (conductivity)
Process temperature below the lowest
value in the table
‣
Extend table.
951
Process
temperature
F
On
On
C
Concentration table (conductivity)
Process temperature above the highest
value in the table
‣
Extend table.
952
Conductivity low
F
On
On
C
Concentration table (conductivity)
Process conductivity below the lowest
value in the table
‣
Extend table.
953
Conductivity high
F
On
On
C
Concentration table (conductivity)
Process conductivity above the highest
value in the table
‣
Extend table.
954
Concentration low
F
On
On
C
Concentration table (conductivity)
Process concentration below the lowest
value in the table
‣
Extend table.
955
Concentration high F
On
On
C
Concentration table (conductivity)
Process concentration above the highest
value in the table
‣
Extend table.
983
Sensor ISE check
F
On
On
I
Electrode or membrane defective
1.
Check or replace electrode.
2.
Check or replace membrane cap.
984
Process temp. high S
On
On
I
Temperature outside specification
1.
Check process temperature.
2.
Check electrode system.
985
Sensor Interface
F
On
On
I
Sensor interface error
1.
Check connector.
2.
Check or replace cable.
987
Calibration
required
M
On
On
I, DI
Electrode replacement
‣
Calibrate sensor.
1)
Status signal
2)
Diagnostic message
3)
Failure current
12.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status.
Furthermore, various service functions are available.