Liquiline CM44P
Diagnostics and troubleshooting
Hauser
155
No.
Message
Factory settings
Sensor
type
Tests or remedial action
S
D
F
556
Temperature low
S
On
On
C
Process temperature above/below
compensation table
• Process value outside specification
• Table incomplete
‣
Extend table.
557
Temperature high
S
On
On
C
558
Conductivity low
S
On
On
C
Process conductivity above/below
compensation table
• Process value outside specification
• Table incomplete
‣
Extend table.
559
Conductivity high
S
On
On
C
560
Conduc.
compensation
S
On
On
C
Conductivity compensation above/below
compensation table
• Process value outside specification
• Table incomplete
‣
Extend table.
561
Conduc.
compensation
S
On
On
C
720
Membrane change
M
On
Off
I
Membrane cap must be replaced
1.
Replace membrane cap.
2.
Reset timer.
722
Sensor reference
F
On
On
P
Alarm: Reference membrane impedance
too low.
1.
Check or replace sensor.
2.
Check/correct reference limit
value.
723
Sensor reference
M
On
Off
I
Warning: Reference membrane
impedance too low.
Can continue measuring until the alarm
occurs.
1.
Check or replace sensor.
2.
Check/correct reference limit
value.
724
Sensor reference
F
On
On
I
Alarm: Reference membrane impedance
too high.
1.
Check or replace sensor.
2.
Check/correct reference limit
value.
725
Sensor reference
M
On
Off
I
Warning: Reference membrane
impedance too high.
Can continue measuring until the alarm
occurs.
1.
Check or replace sensor.
2.
Check/correct reference limit
value.
734
Calibration quality
M
On
Off
O (opt.)
Warning: The calibration quality index
indicates a substantial change since the
last calibration.
Measurement can still take place.
1.
Repeat calibration.
2.
Check the sensor and replace if
necessary.