Operation
Liquiline CM44P
92
Hauser
Outputting the status of a cleaning function
Function = Cleaning
Function
Options
Info
Assignments
Selection
• None
• Depends on the type of
cleaning
Factory setting
None
Here you can specify how a cleaning function
should be displayed for the relay.
You have the following options depending on the
cleaning program that is selected (
Menu
/
Setup
/
Additional functions
/
Cleaning
:
•
Cleaning type
=
Standard clean
Cleaning 1 - Water
,
Cleaning 2 - Water
,
Cleaning 3 - Water
,
Cleaning 4 - Water
•
Cleaning type
=
Chemoclean
Cleaning 1 - Water
,
Cleaning 1 - Cleaner
,
Cleaning 2 - Water
,
Cleaning 2 - Cleaner
,
Cleaning 3 - Water
,
Cleaning 3 - Cleaner
,
Cleaning 4 - Water
,
Cleaning 4 - Cleaner
•
Cleaning type
=
Chemoclean Plus
4x
Cleaning 1 - %0V
, 4x
Cleaning 2 - %0V
Hold behavior
Selection
• Freeze last value
• Fixed value
• Ignore
Factory setting
Ignore
Freeze last value
The device freezes the last measured value.
Fixed value
You define a fixed measured value that is output
at the output.
Ignore
A hold has no effect.
1)
%0V is variable text which you can assign in
Menu
/
Setup
/
Additional functions
/
Cleaning
/
Chemoclean
Plus
/
Output label 1 ... 4
.
Formula
Function = Controller
Function
Options
Info
Operating mode
Selection
• PWM
• PFM
Factory setting
PWM
PWM=pulse width modulation
PFM=pulse frequency modulation→ 90
Source of data
Selection
• List of available
formulas
• Maximum of 8 formulas
Mathematical functions of type Formula must be
available.
‣
Select the formula which is to serve as a data
source.
Range lower value
0 to 9999
Range upper value
Range lower value
to
9999
Cycle duration
Operating mode
=
PWM
Shortest turn-on time
to
999.0 s
Factory setting
10.0 s
‣
Specify the cycle duration within which the
duty cycle should be varied (only PWM).
The settings for
Cycle duration
and
Shortest turn-on time
affect one another. The following applies:
Cycle duration
≥
Shortest turn-on time
.
Shortest turn-on time
Operating mode
=
PWM
0.3 s to
Cycle duration
Factory setting
0.3 s
Pulses that are shorter than this limit value are
not output in order to conserve the actuator.