Liquiline CM44P
Operation
Hauser
71
HART bus address
The list of diagnostic messages displayed depends on the path selected. There are device-
specific messages, and messages that depend on what sensor is connected.
Menu/Setup/General settings/Extended setup/HART
Function
Options
Info
Bus address
0 to 63
Factory setting
0
You can change the device address to integrate
several HART devices in a single network
(Multidrop mode).
If you reset the device to the factory settings (
Diagnostics
/
Reset
/
Factory default
),
the bus address is not reset. Your setting is retained.
PROFIBUS DP
Menu/Setup/General settings/Extended setup/PROFIBUS
Function
Options
Info
Enable
Selection
• Off
• On
Factory setting
On
You can switch off communication at this point.
The software can then only be accessed via local
operation.
Termination
Read only
If the device is the last in the bus, you can
terminate via the hardware.
→ 47
Bus address
1 to 125
If you have addressed the bus via hardware (DIP
switches on the module, → 47), you can only
read the address here.
If an invalid address is set via the hardware, you
have to assign a valid address for your device
either here or via the bus.
Ident number
Selection
• Automatic
• PA-Profile 3.02 (9760)
• Liquiline CM44x (155D)
• Manufacturer specific
Factory setting
Automatic
Modbus
Menu/Setup/General settings/Extended setup/Modbus
Function
Options
Info
Enable
Selection
• Off
• On
Factory setting
On
You can switch off communication at this point.
The software can then only be accessed via local
operation.
Termination
Read only
If the device is the last in the bus, you can
terminate via the hardware.
→ 47