Diagnostics and troubleshooting
Liquiline CM44P
162
Hauser
DIAG/Logbooks
Function
Options
Info
All data logbooks
Data logbook 1 ... 8
All event logbooks
Calibration logbook
Diagnostic logbook
Configuration logbook
HW version logbook
Version logbook
Action,
commences as soon as the
option is selected
Use this function to save the logbook to an SD
card.
‣
Save the logbook in the preferred file
format. You can then open the saved CSV
file on the PC in MS-Excel, for example, and
make further edits here. You can import the
FDM files into Fieldcare and archive them so
they are tamper-proof.
The file name consists of
Logbook ident
(
Menu
/
Setup
/
General settings
/
Logbooks
), an abbreviation for
the logbook and a time stamp.
1)
CSV files use international number formats and separators. Therefore they must be imported into MS
Excel as external data with the correct format settings. If you double-click the file to open it, the data are
only displayed correctly if MS Excel is installed with the US country setting.
12.10 Simulation
You can simulate values at inputs and outputs for testing purposes:
• Current values at current outputs
• Measured values at inputs
• Relay contact opening or closing
Only current values are simulated. Via the simulation function, it is not possible to
calculate the totalized value for the flow or rainfall.
DIAG/Simulation
Function
Options
Info
Current output x:y
Simulation of an output current
This menu appears once for each current output.
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the value at the current output,
this is indicated on the display by a simulation
icon in front of the current value.
Current
2.4 to 23.0 mA
Factory setting
4 mA
‣
Set the desired simulation value.
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation
Selection
• Off
• On
Factory setting
Off
If you simulate the relay state, this is indicated
on the display by a simulation icon in front of
the relay display.
State
Selection
• Low
• High
Factory setting
Low
‣
Set the desired simulation value.
The relay switches in accordance with your
setting when you switch on the simulation. In
the measured value display you see
On
(=
Low
)
or
Off
(=
High
) for the simulated relay state.