Operation
Liquiline CM44P
100
Hauser
Menu/Setup/Outputs/Binary output x:y
Function
Options
Info
Source of data
Selection
• None
• Sensor inputs
• Binary inputs
• Controller
• Fieldbus signals
• Mathematical functions
Factory setting
None
Source, whose value should be read out as a
frequency over the binary output.
Measuring value
Source of data
≠ Controller
Selection
Depends on:
Source of data
‣
Choose the measured value which should be
output as the frequency via the binary
output.
Actuator type
Source of data
= Controller
Selection
• None
• Bipolar
• U
• Unipolar-
Factory setting
None
‣
Determines which controller component the
connected actuator, e. g. the dosing pump,
should trigger.
Bipolar
"Split range"
U
Part of the manipulated variable that the
controller uses to increase the process value
Unipolar-
For connected actuators that decrease the
controlled variable
Hold behavior
Selection
• Freeze
• Fixed value
• None
Factory setting
None
Freeze
The device freezes the last value.
Fixed value
You define a fixed current value that is
transmitted at the output.
None
A hold does not affect this output.
Hold value
Hold behavior
=
Fixed value
0 to 100 %
Factory setting
0 %
Error behavior
Selection
• Freeze
• Fixed value
Factory setting
Fixed value
Freeze
The device freezes the last value.
Fixed value
You define a fixed current value that is
transmitted at the output.
Error value
Error behavior
=
Fixed value
0 to 100 %
Factory setting
0 %
1)
x:y = slot no. : input number
10.6 Additional functions
10.6.1 Limit switches
There are different ways to configure a limit switch:
• Assigning a switch-on and switch-off point
• Assigning a switch-on and switch-off delay for a relay
• Setting an alarm threshold and also outputting an error message
• Starting a cleaning function