Operation
Liquiline CM442R/CM444R/CM448R
82
Hauser
10.5.3 Configuration of binary outputs
Meas. value format
Options
• #
• #.#
• #.##
• #.###
Factory setting
#.##
Specify the number of decimal places.
Input mode
Options
• Frequency
• Parameter
Factory setting
Frequency
Frequency
Display in Hz in the measuring menu
Parameter
You subsequently determine the parameter name and
unit. These are then displayed in the measuring menu.
Parameter name
Input mode= "Parameter"
Free text
Define a name for the parameter, e.g. "flow rate."
Unit of measure
Input mode= "Parameter"
Free text
Determine the unit in which your parameter should be
displayed, e.g." l/min."
Lower range value
Input mode= "Parameter"
-2000.00 to 0.00
Factory setting
0.00
Your previously defined unit is additionally displayed.
The start of measuring range corresponds to a frequency
of 0 Hz.
Upper range value
Input mode= "Parameter"
0.00 to 10000.00
Factory setting
0.00
The end of measuring range corresponds to the
maximum frequency defined above. The displayed unit
is the one predefined by you.
Damping
0 to 600 s
Factory setting
0 s
The damping generates a floating average curve of the
measured value over the specified time.
1)
x:y = slot no. : input number
Path: Menu/Setup/Outputs/Binary output x:y
1)
Function
Options
Info
Binary output
Options
• Off
• On
Factory setting
On
Switching the output on/off
Path: Menu/Setup/Inputs/Binary input x:y
1)
Function
Options
Info