Liquiline CM442R/CM444R/CM448R
Diagnostics and troubleshooting
Hauser
127
11.11 Device test
Alarm relay
Relay x:y
Simulation of a relay state
This menu appears once for each relay.
Simulation
Options
• On
• Off
Factory setting
Off
If you simulate the relay state, this is indicated on the
display by a simulation icon in front of the relay display.
State
Options
• Low
• High
Factory setting
Low
Set the desired state.
The relay switches in accordance with your setting when
you switch on the simulation. The display shows "On"
(="Low") or "Off" (="High") for the simulated relay state.
Meas. inputs
Simulation of a measured value
This menu appears once for each measuring input.
Channel : parameter
Sim. main value
Options
• On
• Off
Factory setting
Off
If you simulate the measured value, this is indicated on
the display by a simulation icon in front of the measured
value.
Main value
Depends on the sensor
Factory setting
Depends on the sensor
Set the desired simulation value.
Sim. temperature
Options
• On
• Off
Factory setting
Off
If you simulate the temperature measured value, this is
indicated on the display by a simulation icon in front of
the temperature.
Temperature
-50.0 to +250.0 °C
(-58.0 to 482.0 °F)
Factory setting
20.0 °C (68.0 °F)
Set the desired simulation value.
Diagnostics/Systemtest
Function
Options
Info
Power supply
Read only
• Digital Supply 1: 1.2V
• Digital Supply 2: 3.3V
• Analog Supply: 12.5V
• Sensor Supply: 24V
• Temperature
Detailed list of power supply to instrument.
The actual values can vary without a malfunction
being present.
Diagnostics/Simulation
Function
Options
Info