Liquiline CM442R/CM444R/CM448R
Operation
Hauser
107
Redundancy
Use this function to monitor two or three sensors with redundant measurements. The
arithmetic average of the two closest measured values is calculated and output as the
redundancy value.
rH value
To calculate the rH value, a pH sensor and an ORP sensor must be connected. It is irrelevant
whether you are using a pH glass sensor, an ISFET sensor or the pH electrode of an ISE sensor.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/Redundancy
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function
Y1
The options depend on the
sensors connected
You can select different types of sensor that output the
same measured value.
Example for temperature redundancy:
You have a pH sensor and an oxygen sensor at inputs 1
and 2. Select both as "Y1" and "Y2". For "Measured value"
select the temperature.
Measured value
Y2
Measured value
Y3 (optional)
Measured value
Deviation control
Options
• Off
• On
Factory setting
Off
You can monitor the redundancy. Specify an absolute
limit value that may not be exceeded.
Deviation limit
Depends on the selected
measured value
Redundancy
Read only
View this value in a user-defined measuring screen or
output the value via the current output.
Path: Menu/Setup/Additional functions/Mathematical functions/MF 1 to 6/Mode/rH calculation
Function
Options
Info
Calculation
Options
• Off
• On
Factory setting
Off
On/off switch for the function