Operation
Liquiline CM442R/CM444R/CM448R
60
Hauser
Logbook name
Customized text, 20
characters
You can change the name here again.
Scan time
00:00:01 to 01:00:00
Factory setting
00:01:00
As above
Minimum interval between two entries
Format: HH:MM:SS
Data logbook
Options
• Off
• Ring buffer
• Fill up buffer
Factory setting
Off
Ring buffer
If the memory is full, the most recent entry
automatically overwrites the oldest entry.
Fill up buffer
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a
corresponding diagnostic message. The memory then
has to be cleared manually.
Line plotter
Menu to define the graphic display
Axes
Options
• Off
• On
Factory setting
On
Should the axes (x, y) be displayed (On) or not (Off)?
Orientation
Options
• Horizontal
• Vertical
Factory setting
Horizontal
You can choose whether the value curves should be
displayed from left to right ("Horizontal") or from top to
bottom ("Vertical").
If you want to display two data logbooks simultaneously,
make sure that both logbooks have the same settings
here.
X-Description
Options
• Off
• On
Factory setting
On
Decide whether a description should be displayed for the
axes and whether gridlines should be shown. In
addition, you can also decide whether pitches should be
displayed.
Y-Description
Grids
Pitches
X Pitch/Grid distance
10 to 50%
Factory setting
10 %
Specify the pitches.
Y Pitch/Grid distance
Remove
Action
This action removes the data logbook. Any data that
have not been saved are lost.
Path: Menu/Setup/General settings/Logbooks
Function
Options
Info