Diagnostics and troubleshooting
Liquiline CM442R/CM444R/CM448R
120
Hauser
374 Sensor check
F
On
Off
No measurement signal from sensor
-> Check the sensor connection
-> Check sensor, replace if necessary
401 Factory reset
F
On
On
Factory reset is performed
406 Param. active
C
Off
Off
→
Wait for con
fi
guration to be
fi
nished
407 Diag. active
C
Off
Off
→
Wait for maintenance to be
fi
nished
412 Writing backup
F
On
Off
→
Wait for the write process to be
fi
nished
413 Reading backup
F
On
Off
→
Wait
460 Output below limit
S
On
Off
Reasons
• Sensor in air
• Air pockets in assembly
• Sensor fouled
• Incorrect flow to sensor
Measures
1. Check sensor installation
2. Clean the sensor
3. Adjust assignment of current outputs
461 Output above limit
S
On
Off
462 Output Deviation
S
On
Off
When the current output is read back, the value
deviates from the target value displayed.
Possible reasons: Current load outside
specification, short-circuit or open current loop,
module defective
1. Check installation of current loop
2. Check module
3. Contact the Service Department
502 No text catalog
F
On
On
→
Contact the Service Department
503 Language change
M
On
Off
Language change failed
→
Contact the Service Department
530 Logbook at 80%
M
On
Off
1. Save the logbook to the SD card and then
delete the logbook in the device
2. Set memory to ring memory
3. Deactivate logbook
531 Logbook full
M
On
Off
532 License error
M
On
Off
→
Contact the Service Department
540 Parameter save fail
M
On
Off
Con
fi
guration saving has failed,
→
repeat
541 Parameter load ok
M
On
Off
Configuration successfully loaded
542 Parameter load fail
M
On
Off
Con
fi
guration loading has failed,
→
repeat
543 Parameter load abort
M
On
Off
Configuration loading aborted
544 Parameter reset ok
M
On
Off
Factory default successful
910 Limit switch
S
On
Off
Limit switch activated
No.
Message
Factory settings
Tests or remedial measures
Cat. Diag.
on/off
Error
current
Содержание Liquiline CM442R
Страница 2: ......
Страница 170: ...Liquiline CM442R CM444R CM448R 170 ...
Страница 171: ...Liquiline CM442R CM444R CM448R Endress Hauser ...