Repair
Source container FQG66
40
Hauser
10.4.4 Pre-return inspection
Company
Name
Address
Name of inspector and role
Container
FQG6_-_____________
Radiation source
Isotope
137
Cs
60
Co
Source serial number
Nominal activity (MBq / GBq)
Date of manufacture
Checks
Result
Wipe test report not older than 3 months is enclosed with the return shipment
documents
A copy of the manufacturer's certificate of the source is enclosed with the return
shipment documents
No significant signs of corrosion which could put the safe storage of the source at
risk.
No signs of serious damage on the container from fire, falls or collisions.
"EIN/ON" and "AUS/OFF" mechanism works according to the Operating
Instructions
The container is in the "AUS/OFF" position and secured with a padlock
The transport index has been determined
The container is labeled according to international regulations concerning the
transportation of hazardous materials (ADR/RID, DGR/IATA)
Date
Signature
10.5 Disposing of the measuring device
L
WARNING
Danger to personnel and environment from fluids that are hazardous to health.
‣
Ensure that the measuring device and all cavities are free of fluid residues that are
hazardous to health or the environment, e.g. substances that have permeated into
crevices or diffused through plastic.
Observe the following notes during disposal:
‣
Observe valid federal/national regulations.