Installation
Source container FQG66
18
Hauser
2
1
1
4
0
°
4
0
°
A0023679
1
FQG66: Order code 240 "Angle of emission", option 5 "40 degrees"
2
Gammapilot
5.2.2
Orientation for point level detection
For point level detection, the radiation source container is mounted at the same height as
the detector.
1
2
5°
A0023681
1
FQG66: Order code 240 "Angle of emission", option 1 "5 degrees, horizontal"
2
Gammapilot
• Keep the distance between the FQG66 and the vessel wall to a minimum!
• Secure any intermediate area between the source container and wall by grip
protection, if necessary!