Source container FQG66
Commissioning
Hauser
27
4 →
5 →
6
4
5
3
1
2
A0027126
A0027127
A0027128
‣
Pull out the source holder rod
(5) as far as the "EIN/ON"
position. Reinsert the lock pin
(4) into the bore hole of the
guide tube and push it in as far
as it will go.
‣
Fit the cover (3).
‣
Secure the cover with the
screws (2) (6mm across flats)
and fit the padlock (1) back on.
7.3.2
Pneumatic version
The measuring device must first be connected to the compressed air supply before the
pneumatic version can be put into operation .
L
CAUTION
Risk of injury when cover is open
‣
Make sure that the pneumatic drive is unpressurized for the entire time the cover is
removed!
The device can be switched ON and OFF with the pneumatic controller.
1 →
2 →
3 →
1
2
3
4
A0027129
A0027131
A0027132
‣
Remove padlock (1). Turn the
screws (2) (6mm across flats)
on the cover three to four
times to release them.
‣
Remove the cover (3).
‣
Remove the lock pin (4) from
the guide tube of the source
holder rod and insert it into
the bore hole ("parking
position") provided.