8.2.4
What to do in the event of corrosion
If there are clear signs of corrosion at the source container, the local dose rate around the
device must be measured. If the value is significantly above the normal operation levels,
cordon off the area and notify the radiation safety officer responsible.
L
CAUTION
What to do if the radiation source container is damaged
‣
Corroded source containers must be replaced immediately
‣
Only use a genuine spare part to replace a damaged padlock
8.3
Measuring and test equipment
Dosimeter to check the control zone
8.4
Hauser services
Hauser offers a wide variety of services for maintenance such as recalibration,
maintenance service or device tests.
Your Hauser Sales Center can provide detailed information on the services.