Deltapilot S
14
Hauser
Output
Output signal
• 0.2 to 1.2 mA analog signal, 3-wire
• 200 to 1200 Hz PFM signal, 2-wire
– f
0
= 200 Hz ± 5 Hz
– 100 mbar measuring range: f
0
= 200 Hz ± 10 Hz
• 4 to 20 mA with superimposed communication protocol HART, 2-wire
• Digital communication signal PROFIBUS PA (Profile 3.0), 2-wire
• Digital communication signal FOUNDATION Fieldbus, 2-wire
Signal on alarm
• 0.2 to 1.2 mA analog:
≥
1.5 mA
• PFM
≤
20 Hz
• 4 to 20 mA HART
Options:
– Min. = 3.6 mA; onsite display: 0
– Max. = 22 mA (factory setting); onsite display: 1
– Hold: last measured value is held; onsite display: 2
• PROFIBUS PA:
– Options configurable in the Analog Input Block for the output signal:
Last Valid Out Value, FSafe Value (factory setting), Status bad
– Options configurable in the Transducer Block for the "Measured value" parameter and the onsite display:
Min. (onsite display: –19999), Max. (onsite display: +19999), Hold: last measured value is held
• FOUNDATION Fieldbus:
– Output signal: last measured value is held; status change from "Uncertain" or "Bad"
– Options configurable in the Transducer Block for the "Measured value" parameter and the onsite display:
Min. (onsite display: –19999), Max. (onsite display: +19999), Hold: last measured value is held
Load
Sensitivity, PFM –FEB17(P)
Damping
• 0 to 99 s configurable via the FHB20 display and operating module, PC with operating program or handheld
terminal
• Factory setting: 0 s
0.1 to 1.2 mA analog
PFM
4 to 20 mA HART
Max. 25
Ω
/core
Max. 25
Ω
/core
P01-DB5xxxxx-05-xx-xx-xx-002
820
0
30
11.5
Ub
[V]
Rb
[ ]
Ω
Measuring range
Resolution
Measuring range
Resolution
0 to 100 mbar
10 Hz/mbar
–100 to 100 mbar
5 Hz/mbar
0 to 400 mbar
2.5 Hz/mbar
–400 to 400 mbar
1.25 Hz/mbar
0 to 1200 mbar
0.833 Hz/mbar
–900 to 1200 mbar
0.476 Hz/mbar
0 to 4000 mbar
0.25 Hz/mbar
–900 to 4000 mbar
0.204 Hz/mbar
0 to 10000 mbar
0.1 Hz/mbar
–900 to 10000 mbar
0.092 Hz/mbar