Cubemass DCI
Hauser
5
Input
Measured variable
• Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Volume flow (calculated using mass flow and density)
• Fluid density (proportional to the resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
Measuring range
Measuring ranges for liquids
Operable flow range
1:100
Input signal
Status input (auxiliary input)
U = 3 to 30 V DC, R
i
= 5 k
Ω
, galvanically isolated.
Switching level: 3 to 30 V DC, polarity-independent.
Configurable for: totalizer reset, positive zero return, error message reset, start zero point adjustment.
DN
Range for full scale values (liquids)
g
min(F)
to
g
max(F)
[mm]
[inch]
[kg/h]
[lb/min]
1
1/24"
0 to 20
0 to 0.75
2
1/12"
0 to 100
0 to 3.7
4
1/8"
0 to 450
0 to 16.5
6
¼"
0 to 1000
0 to 37