Cubemass DCI
10
Hauser
Performance characteristics
Reference operating
conditions
• Error limits following ISO/DIS 11631
• Water, typically 20 to 30 °C (68 to 86 °F); 2 to 4 bar (30 to 60 psi)
• Data as per the calibration report ±5 °C (±9 °F) and ±2 bar (±30 psi)
• Data on the measured error based on accredited calibration rigs traced back to ISO 17025
Maximum measured error
The values indicated refer to the pulse/frequency output.
The additional measured error for the current output is typically ±5
μ
A. Basis for calculations
o.r. = of reading
Mass flow and volume flow (liquids)
• ±0.10% o.r. (mass flow)
• ±0.10% o.r. (volume flow)
Density (liquids)
• ±0.001 g/cc (after field density calibration or under reference conditions)
• ±0.002 g/cc (special density calibration (optional)
Calibration range: 0.0 to 2.0 g/cc, 5 to 80 °C (41 to 176 °F)
Application range: 0.0 to 5.0 g/cc, –50 to 200 °C (–58 to 392 °F)
• ±0.02 g/cc (standard calibration)
Temperature
±0.5 °C ± 0.005 · T °C
(±1.0 °F ± 0.003 · (T – 32) °F)
T = Fluid temperature
Zero point stability
Example of maximum measured error
A0011691
Max. measured error in % o.r. (example: Cubemass DCI, DN 1)
DN
Max. full scale value
Zero point stability
[mm]
[inch]
[kg/h]
[lb/min]
[kg/h]
[lb/min]
1
1/24"
0 to 20
0 to 0.75
0.0008
0.00003
2
1/12"
0 to 100
0 to 3.7
0.002
0.00007
4
1/8"
0 to 450
0 to 16.5
0.014
0.0005
6
¼"
0 to 1000
0 to 37
0.02
0.0007
lb/min
0
0.8
0.2
1.0
0.4
0.6
[%]
kg/h
0
0.5
0.2
1.0
0.1
0.3
0.4
0.6
0.7
0.8
0.9
0
25
5
10
15
20