Rupture element/disk
The sensor housing is optionally available with an integrated rupture element.
#
Warning!
• Make sure that the function of the rupture element is not impeded by the installation. The triggering pressure
in the housing is indicated on the information notice. Take suitable measures to ensure that no damage can
occur if the rupture disk is tripped and personal injury is ruled out.
Triggering pressure in the housing 10 to 15 bar (145 to 217 psi)
• Please note that if a rupture disk is used, the housing can no longer assume a secondary containment
function.
• It is not permitted to open the connections or remove the rupture disk.
"
Caution!
The existing connection nozzles are not designed for a purge or pressure monitoring function.
!
Note!
• The transportation guard on the rupture disk must be removed prior to commissioning.
• Comply with the information on information notices.
A0012344
1 = Rupture element, 2 = Transportation guard
A0011967
Additional sign regarding the position of the rupture disk (RUPTURE DISK)
A
Sensor housing with rupture element (incl. rupture disk)
È
defined fluid exit
A
Sensor housing with rupture disk
È
undefined fluid exit
!
Note!
The location of the rupture disk with undefined fluid exit is covered over by an adhesive label. If the rupture
disk is tripped, the adhesive label is destroyed, meaning that a visual check is possible.
2
1
RUPTURE DISK
12.5BAR+/-10%@80°C
i
RUPTURE DISK
12.5BAR+/-20%@20°C
i
A
B