Cerabar PMC51B HART

Commissioning

Hauser

21

3. When the LED lights up briefly, the pressure present has been accepted for the lower

range value.

Setting the upper range value (pressure or scaled variable)

1. The desired pressure for the upper range value is present at the device.
2. Press the "Span" key for at least 3 s.
3. When the LED lights up briefly, the pressure present has been accepted for the upper

range value.

4. Does the LED on the electronic insert not light up?

 Applied pressure has not been accepted for upper range value.

Wet calibration is not possible if the 

Scaled variable

 option has been selected in

the 

Assign PV

 parameter and the 

Table

 option has been selected in the 

Scaled

variable transfer function

 parameter.

Checking the settings (pressure or scaled variable)

1. Press the "Zero" key briefly in order to display the lower range value.
2. Press the "Span" key briefly in order to display the upper range value.
3. Press the "Zero" and "Span" keys simultaneously in order to display the calibration offset.

Resetting the device

Press and hold the "Zero" and "Span" keys simultaneously for at least 12 s.

8.5.2  

Commissioning with the commissioning wizard

In FieldCare, DeviceCare

 1)

, SmartBlue and on the display, the 

Commissioning

 wizard is

available to guide the user through the initial commissioning steps. Commissioning is also

possible via AMS or PDM.

1. Connect the device with FieldCare or DeviceCare.
2. Open the device in FieldCare or DeviceCare.

 The dashboard (homepage) of the device is displayed:

3. In the 

Guidance

 menu, click the 

Commissioning

 wizard to open the wizard.

4. Enter the appropriate value in each parameter or select the appropriate option. These

values are written directly to the device.

5. Click "Next" to go to the next page.
6. Once all the pages have been completed, click "End" to close the 

Commissioning

 wizard.

If the 

Commissioning

 wizard is cancelled before all the necessary parameters have been

configured, the device may be in an undefined state. In such situations, it is advisable to

reset the device to the factory default settings.

1)

DeviceCare is available for download at www.software-products.endress.com. You must register in the Endress

+Hauser software portal to download the product.

Содержание Cerabar PMC51B

Страница 1: ...hey are not a substitute for the Operating Instructions pertaining to the device Detailed information about the device can be found in the Operating Instructions and the other documentation Available...

Страница 2: ...dress com deviceviewer Endress Hauser Operations App XXXXXXXXXXXX XXXXX XXXXXX XXX XXXX XX Serial number 1 3 2 A0023555 2 About this document 2 1 Document function The Brief Operating Instructions con...

Страница 3: ...tion on procedures and other facts which do not result in personal injury 2 2 2 Electrical symbols Ground connection Terminal for connection to the grounding system 2 2 3 Symbols for certain types of...

Страница 4: ...nnel The personnel for installation commissioning diagnostics and maintenance must fulfill the following requirements Trained qualified specialists must have a relevant qualification for this specific...

Страница 5: ...ure continued operational safety and reliability Carry out repairs on the device only if they are expressly permitted Observe federal national regulations pertaining to the repair of an electrical dev...

Страница 6: ...ce offers specific functions to support protective measures by the operator These functions can be configured by the user and guarantee greater in operation safety if used correctly An overview of the...

Страница 7: ...e installation Always firmly tighten the housing cover and the cable entries 1 Counter tighten the cable entries 2 Tighten the coupling nut 5 1 2 Installation instructions The devices are mounted acco...

Страница 8: ...tighten the NPT thread to the required depth according to the standard For the following process connections a tightening torque of max 40 Nm 29 50 lbf ft is specified Thread ISO228 G with flush membr...

Страница 9: ...pping point so that any condensate can flow into the process 5 2 2 Pressure measurement in vapors A water pocket pipe reduces the temperature to almost the ambient temperature The defined water column...

Страница 10: ...OTICE Thread and housing cover damaged from dirt and fouling Remove fouling e g sand on the thread of the cover and housing If you continue to encounter resistance when closing the cover check the thr...

Страница 11: ...thread The thread of the electronics and connection compartment is coated with lubricant varnish Avoid additional lubrication 6 2 1 Supply voltage Ex d Ex e non Ex supply voltage 10 5 to 35 VDC Ex i...

Страница 12: ...tched on 6 2 2 Terminals Supply voltage and internal ground terminal 0 5 to 2 5 mm2 20 to 14 AWG External ground terminal 0 5 to 4 mm2 20 to 12 AWG 6 2 3 Cable specification Protective ground or groun...

Страница 13: ...deration Maximum 6 V with a communication resistor of 250 6 2 5 Overvoltage protection Devices without optional overvoltage protection Equipment from Endress Hauser fulfills the requirements of the pr...

Страница 14: ...with IEC EN 61010 The cables must be adequately insulated with due consideration given to the supply voltage and the overvoltage category The connecting cables must offer adequate temperature stabilit...

Страница 15: ...onnection compartment 1 Positive terminal 2 Negative terminal 3 Internal ground terminal 6 2 8 Cable entries 1 2 A0045413 1 Cable entry 2 Dummy plug The type of cable entry depends on the device versi...

Страница 16: ...ation of moisture into the device Devices with valve plug 2 1 A B BU BN GNYE 2 1 3 A0023097 3 BN brown BU blue GNYE green yellow A Electrical connection for devices with valve connector B View of the...

Страница 17: ...sport protection IP22 TYPE 2 Cable 5 m IP66 68 TYPE 4X 6P pressure compensation via cable Valve plug ISO4400 M16 IP65 TYPE 4X HAN7D plug 90 degrees IP65 NEMA Type 4X M12 plug When housing is closed an...

Страница 18: ...switch for locking and unlocking the device The setting of the DIP switches has priority over the settings made via other operation methods e g FieldCare DeviceCare 7 2 Access to the operating menu v...

Страница 19: ...plate LWARNING The settings of the current output are relevant for safety This situation can result in product overflow The setting of the current output depends on the setting in the Assign PV parame...

Страница 20: ...ng tool See the description of the operating tool 8 5 Configuring the measuring device 8 5 1 Commissioning with keys on the electronic insert The following functions are possible via the keys on the e...

Страница 21: ...ion offset Resetting the device Press and hold the Zero and Span keys simultaneously for at least 12 s 8 5 2 Commissioning with the commissioning wizard In FieldCare DeviceCare 1 SmartBlue and on the...

Страница 22: ...easured value when the vessel is empty or partly filled the measured value is not zero Perform a position adjustment if necessary In the Assign PV parameter the Pressure option must be selected factor...

Страница 23: ...ressure shifts in the measured value when the vessel is empty or partly filled the measured value is not zero Perform position adjustment if necessary B 450 mbar 1000 l A 50 mbar 0 l V l 1000 0 50 p m...

Страница 24: ...ure value 2 4 Enter the volume value for the upper calibration point via the Scaled variable value 2 parameter 1000 l 264 gal Menu path Application Sensor Scaled variable Scaled variable value 2 Resul...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ...www addresses endress com 71582834 71582834...

Отзывы: