Cerabar PMC51B HART
Electrical connection
Hauser
13
6.2.4
4-20 mA HART
2
3
4
Y
+
+
-
-
I
1
mA
A0028908
1
Block diagram of HART connection
1
Device with HART communication
2
HART communication resistor
3
Power supply
4
Multimeter
The HART communication resistor of 250 Ω in the signal line is always necessary in the
case of a low-impedance power supply.
Take the voltage drop into consideration:
Maximum 6 V with a communication resistor of 250 Ω
6.2.5
Overvoltage protection
Devices without optional overvoltage protection
Equipment from Hauser fulfills the requirements of the product standard IEC / DIN
EN 61326-1 (Table 2 Industrial Environment).
Depending on the type of port (DC power supply, input/output port) different testing levels
according to IEC / DIN EN 61326-1 against transient overvoltages (Surge) are applied (IEC /
DIN EN 61000-4-5 Surge):
Test level on DC power ports and input / output ports is 1000 V line to earth
Overvoltage category
Overvoltage category II