Trouble-shooting
ASP Inline
46
Hauser
9.3
Process errors without messages
9.4
Spare parts
When ordering accessories or spare parts, please state the serial number of the machine.
9.5
Return
You should take the following measures before sending the device back to Hauser, for
example for repair:
• Always enclose a fully-completed "Dangerous Goods Sheet" with the device. Only then can
Hauser transport, test or repair a returned device.
• If necessary, include special handling regulations with the returned product, e.g. a safety
datasheet in accordance with EN 91/155/EEC.
• Remove all signs of fluids. Pay special attention to the grooves for seals and crevices which could
contain residues. This is particularly important if the substance is hazardous to health, e.g.
flammable, toxic, caustic, carcinogenic, etc.
!
Note!
A
master copy
of the ’Dangerous Goods Sheet’ can be found at the end of these Operating
Instructions.
"
Caution!
• Do not return any devices if you cannot be entirely sure that you have completely removed all
traces of hazardous substances, e.g. any media which have penetrated crevices or diffused
through plastic.
• Costs incurred for waste disposal and injury (burns etc.) due to inadequate cleaning will be
charged to the owner-operator.
Error
Cause
Solution
No sample liquid in sampling tank/
sampling device
• Drain hose congested
• Drain hose not connected
• Drain hose with syphon
• Dosing equipment/valve ball
congested
• Rinse drain hose
• Check connection
• Check position of the drain hose
• Clean the ball hole with a round
brush
Leak at the assembly
Shell seals defective
Replace shell seals
Sampler assembly does not trigger
sampling
• Pneumatics not connected
• Control signal incorrectly wired
• Check pneumatic connection
Check compressed air supply
• Check wiring
Order number
Spare part
RPM20X-DA
5 ml dosing ball with stainless steel lever
RPM20X-DB
50 ml upper dosing ball with stainless steel lever
RPM20X-DC
50 ml lower dosing ball with stainless steel lever
51005590
Body gasket 50 ml ASP Inline pressure sampling
51004746
Body gasket 5 ml ASP Inline pressure sampling
51005591
Gasket for ASP Inline flange
RPM20X-ZA
Pneumatic cylinder for 5 ml version
RPM20X-ZB
Pneumatic cylinder for 50 ml version
RPM20X-VA
5/2-way valve 24 V