ASP Inline
Commissioning
Hauser
43
6.3
Device configuration
When connecting the ASP Inline sampler assembly to the ASP Station 2000 stationary sampler,
please read the ’ASP Station 2000’ Operating Instructions about configuring the sampling
programme (
BA080r09
) - chapter 6.3 to 6.6.
When connecting the ASP Inline sampler assembly to other control units, read the relevant chapter
of the corresponding operating instructions about configuring the sampling programme.
6.4
User-specific applications
For the ASP Inline assembly configuration, enter the following settings in the control unit of the ASP
Station 2000 stationary sampler (see
BA080r09
). Select the options TIME and SHOT-VOL. under
SETUP
Æ
BASIC SETTINGS
Æ
SAMPLING:
Fig. 14:
ASP Inline sampling menu structure (via ASP Station 2000 control unit)
6.4.1
Pre-Shots
The Pre-Shots function can be configured for every standard device.
SETUP -> BASIC SETTINGS -> Sampling -> Pre-Shots -> (None, 1, 2, 3, 4, 5)
!
Note!
For the correct operation of the Pre-Shot function, the ASP Station 2000 has to have a rinsing point
with a drain for the tap.
Prior to every sampling, a configurable number of samples ("shots") are rejected. The tap of the ASP
Station 2000 empties the corresponding number of shots into the drain of the rinsing point prior to
every sampling. The tap then returns to the defined sample bottle for sampling.
By pre-rinsing, any dead volume at the process connection of the assembly can be reduced.
Fig. 15:
"Pre-Shots" sampling function, ASP Inline (via ASP Station 2000 controller)
6.4.2
Rinsing
The ASP Inline sampler assembly is optionally available with a rinse function with industrial water.
From the ASP Station 2000 with SW version LDB64XA (or higher e.g. LDB65XA), the rinse
Time
Entry for the switch time between filling the sample chamber and draining the sample.
Shot-Vol.
Setting for the sample volume, which is taken per sampling (shot) (Settable value: 5 to max.
50 ml).
Содержание ASP Inline
Страница 1: ...BA149R 09 a3 02 07 51005630 Betriebsanleitung ASP Inline Probensammler Armatur...
Страница 78: ......
Страница 79: ......
Страница 80: ...www endress com worldwide BA149R 09 a3 02 07 51005630 FM SGML6 0 ProMoDo...