57
Symptom
Solution
Liquid flow too low, or instrument tends
to accumulate heat
The pump does not function, see below
•
Remove the pump and check that drive shaft rotates smoothly when activated
•
Check that the silicon tubing is correctly placed in the pump
•
Check that the sterile line is the one recommended by EMS for this unit
•
Check that the sterile line is not damaged or obstructed
•
Check flow adjustment, increase flow to maximum
•
Check that the instrument orifice is not partially clogged up
•
Send the unit to your authorized EMS Service Center
•
Liquid flow too high
Check flow adjustment and decrease it
•
The pump does not work
Check that pump is correctly positioned on support
•
Check that the hand piece is correctly connected to the unit
•
Remove the pump and check if the drive shaft rotates
•
Check that the foot pedal is properly connected to unit
•
Activate the pump with the backup switch on back panel
•
Replace the foot pedal
•
Send the unit to your authorized EMS Service Center
•
The hand piece does not remain on the
magnetic support
Check the seal at the support base
•
Install the magnetic support on a metallic part of the unit
•
The support is demagnetized
•
Replace magnetic support
•
The hand piece cap does not remain
tight
Check for presence of the O-ring seal and replace if necessary
•
Check the thread on handpiece
•
Replace the handpiece
•