19
Transceiver and retro reflector assembly
1.
Attach the Transceiver and Retro reflector to
the air plenum assembly by placing them on
the hinge pins.
2.
Close transceiver & retro and secure in place
with the two hold down latches.
3.
The air-purge blowers should be powered up
at this time to prevent stack particulate from
accumulating in the nipple and air-purge
housing.
Caution:
If installed location has a positive pressure
the air-purge system must be used continuously during installation to prevent process gases from contaminating
optical surfaces or over heating instrument electronics. If the system is shut off for more than momentary
interruptions, the instrument may be damaged. Failure to provide continuous air-purge may void the warranty.
All wiring from the control unit to the transceiver should be completed at this time.
NOTE: THE AIR PLENUM ASSEMBLY FOR BOTH TRANSCEIVER AND RETRO MUST BE
INSTALLED AS BELOW, I.E. THE PINS ON THE LEFT SIDE POINTING UP!
Lift side Upper
Hinge Pin
Left side Lower
Hinge Pin
Содержание 1304
Страница 45: ...44 Section 8 DRAWINGS ...
Страница 46: ...45 EMS 1001 Mounting under 6 STACK ...
Страница 47: ...46 EMS 1002 Mounting over 6 stack ...
Страница 48: ...47 EMS 1024 Air purge Weather cover mech Installation ...
Страница 49: ...48 EMS 1078 PANEL cut out for 1304 control unit ...
Страница 50: ...49 EMS 1079 1304 control unit dimensions ...
Страница 51: ...50 EMS 1077 Terminal identification 1304 control unit ...
Страница 52: ...51 EMS 1131 1304 with recorder system wiring ...
Страница 54: ...53 EMS1101 Sht 1 of 3 Dual Blower system wiring ...
Страница 55: ...54 EMS 1101 Sht 2 of 3 Single Blower system wiring ...
Страница 56: ...55 EMS 1101 Sht 3 of 3 Without Blower system wiring ...