28
Auto cal timer Adjustment
Under the flip down door, press the
EDIT (ESC)
button (the red Edit (ESC) LED will light). The message
display will read QUICK MENU. Press the
ENTER
key, then the
EDIT
key 3 times or scroll down until the
Display reads AUTOCAL CYCL TMR. Press the
ENTER
key twice, the Edit Value Light will flash. Press and
hold the EDIT
or
EDIT
key until the desired set point in hours is displayed, i.e. 24.00 equals 24 hour
intervals. Press the
EDIT
key 4 times to return to normal operation.
Disable internal cal timer
(Version “K” dated 11/28/2007 and up)
Set the cal cycle adjustment for 24.1 hours this will disable the internal clock from initiating a calibration check
cycle. The count down is still displayed but it has no effect. To start a calibration cycle in this mode use the
remote calibration cycle initiate.
Remote calibration cycle initiate
To have an external device cause the monitor to begin calibration cycle, connect a N.O. dry contact to terminal 8L
& 9. A momentary closure will initiate a 3-minute zero and a 3-minute span check cycle. Total time of calibration
check is 6 minutes.
Remote calibration cycle acknowledgment
Cal in progress - By dry contacts on terminal 4U (common) and 5U (NC). This contact remains closed until cal is
completed, i.e. both zero & span.
Содержание 1304
Страница 45: ...44 Section 8 DRAWINGS ...
Страница 46: ...45 EMS 1001 Mounting under 6 STACK ...
Страница 47: ...46 EMS 1002 Mounting over 6 stack ...
Страница 48: ...47 EMS 1024 Air purge Weather cover mech Installation ...
Страница 49: ...48 EMS 1078 PANEL cut out for 1304 control unit ...
Страница 50: ...49 EMS 1079 1304 control unit dimensions ...
Страница 51: ...50 EMS 1077 Terminal identification 1304 control unit ...
Страница 52: ...51 EMS 1131 1304 with recorder system wiring ...
Страница 54: ...53 EMS1101 Sht 1 of 3 Dual Blower system wiring ...
Страница 55: ...54 EMS 1101 Sht 2 of 3 Single Blower system wiring ...
Страница 56: ...55 EMS 1101 Sht 3 of 3 Without Blower system wiring ...