30
Section 5
AJUSTMENTS
Clear on stack zero and span calibration
A clear stack condition must exist to perform this calibration. Power must have been on for no less than 30
minutes. Do not attempt these adjustments in inclement weather. After the cover is removed from the transceiver
normal levels of day light in the area will not effect the calibration.
To complete this procedure the following items are required: EMS Micro-turn
200 on-line test kit (p/n 1232) with a high filter of at least 0.8 O.D.
1.
Swing both the retro and transceiver open and clean the protective
windows (see Figure 4). Return both to the closed position.
2.
Verify alignment, returning beam is centered on the TTL target.
3.
On the transceiver remove the screw below the target viewing window
and pulling the housing straight back until it clears the optical plate.
4.
On the Service Module (Figure 3) make sure the
normal/test
switch is
in the normal position.
5.
NOTE: All adjustments are on the 222-1667 PC Board (see figure 1).
6.
Adjust the 20 turn Zero potentiometer on board 222-1667 marked "PT
-1", CW for an up scale reading >15%, then slowly CCW for 0-1%
opacity.
7.
Install the EMS Micro-turn 200 on-line test
reflector on the transceiver and screw the device to the
transceiver with the mounting screw. With the
thumbwheel adjust the on-line reflector for the same
opacity as in previous step and lock it in place.
8.
Place the highest value filter (for best results at
least a 0.8 O.D.) in the slot provided. Adjust the Span
potentiometer on board 222-1667 marked "PT -3" for
the filters correlated value on the service module
opacity display equals to the correlated value.
See section six (6) for filter correlation
formulas.
9.
Remove the filter and adjust the Zero potentiometer PT-1 for 0-1%.
10.
Insert the High filter again adjust PT-3 for its value, repeat steps until the values come within 0.5 %
Opacity.
11.
Remove the on-line test reflector and replace the transceiver cover and secure the transceiver in place.
You must complete "Zero Reflector Adjustment" procedure next.
Zero reflector adjustment
1.
After clear or off stack zero has been performed the zero reflector needs to be adjusted.
2.
Find and record the zero offset value found in the QUICK MENU under "Zero Cal Value", see
"Zero/Span Calibration Checks" in section 3 for procedure to access the quick menu.
Figure 3
PT-1 Zero
Figure 1
Содержание 1304
Страница 45: ...44 Section 8 DRAWINGS ...
Страница 46: ...45 EMS 1001 Mounting under 6 STACK ...
Страница 47: ...46 EMS 1002 Mounting over 6 stack ...
Страница 48: ...47 EMS 1024 Air purge Weather cover mech Installation ...
Страница 49: ...48 EMS 1078 PANEL cut out for 1304 control unit ...
Страница 50: ...49 EMS 1079 1304 control unit dimensions ...
Страница 51: ...50 EMS 1077 Terminal identification 1304 control unit ...
Страница 52: ...51 EMS 1131 1304 with recorder system wiring ...
Страница 54: ...53 EMS1101 Sht 1 of 3 Dual Blower system wiring ...
Страница 55: ...54 EMS 1101 Sht 2 of 3 Single Blower system wiring ...
Страница 56: ...55 EMS 1101 Sht 3 of 3 Without Blower system wiring ...