Figure 4-3: Sensor with brackets assembled and the protective
cap fitted
A. Bolt heads
3. Hold the brackets in position, using a torque wrench and an allen key
to tighten the fasteners to 14Nm.
4.2
Commissioning the sensor for the magnetic fixture
Commissioning allows the sensor to securely join a designated network and
communicate with a gateway. The IK220 installation kit comes with a
commissioning interface (CC21) and a tablet PC with the Permasense
installation application installed. The CC21 provides an electronic interface
between the ET310 sensor and the tablet PC during commissioning.
The installation application software is used to
1. Provision the WiHART network configuration on the sensor.
2. Monitor the quality of the ultrasonic signal during mechanical
installation.
Both of these steps are required to complete the commissioning.
Note
All the sensors connected to the network and Gateway must have the same
network ID and Join Key.
Procedure
1. Power up the rugged tablet PC and connect the CC21 commissioning
interface to the tablet PC USB port.
January 2022
Quick Start Guide
Quick Start Guide
27