Figure 3-2: Install Tool
3.4
Sensor installation
Procedure
1. Click the
Installation
tab in the installation application software.
2. Click the
Start
button and wait for an ultrasonic waveform to
download from the sensor.
Note
Waveforms are automatically downloaded every 10 seconds. When a
new waveform arrives, the lines briefly become thicker.
3. Check the quality of the waveform.
Need help?
The first one or two reflections must be well defined above the noise
in the signal. Only one reflection is needed to calculate a thickness. If
the signal is poor, move the sensor to a slightly different position.
January 2022
Quick Start Guide
Quick Start Guide
19