March 2016
Reference Manual
00809-0100-4772, Rev FA
175
To configure alert simulation:
1.
Navigate to
Configure
>
Alert Setup
.
2.
Select the
Simulation
tab. The
Simulation
tab displays.
3.
Select the alerts you wish to simulate by checking the check box next to the desired
alert condition.
Note
It is strongly recommended to only select one alert in each category. Selecting multiple
alerts in a category will result in an overabundance of alerts at the control system.