March 2016
Reference Manual
00809-0100-4772, Rev FA
155
C.2.6
Calculating output variables with known input frequency
Use the following equations with a known input frequency for verification of a flow rate.
Select the proper equation depending on whether you are verifying a flow rate or a mass
flow rate.
To verify a volume flow rate
Perform the following calculation to verify a volume flow rate for a given frequency and
K-factor (compensated).
Conditions
Q = flow rate
F = frequency, measured in Hz
K = K-factor (compensated)
C
x
= the unit conversion (see
)
Equation
To verify a standard or normal flow rate
Perform the following calculation to verify a standard or normal flow rate for a given
frequency and K-factor (compensated).
Conditions
Q = flow rate
F = frequency, measured in Hz
K = K-factor (compensated)
C
x
= the unit conversion (see
)
Equation
To verify a mass flow rate
Perform the following calculation to verify a mass flow rate for a given frequency and
K-factor (compensated).
Conditions
M = flow rate
F = frequency, measured in Hz
K = K-factor (compensated)
ρ
= density at operating conditions
C = the unit conversion
Q
F
K
C
x
×
--------------
=
Q
F
Density Ratio
K
C
x
×
--------------------------
×
=