![Emerson Rosemount 5408 Скачать руководство пользователя страница 153](http://html.mh-extra.com/html/emerson/rosemount-5408/rosemount-5408_reference-manual_103670153.webp)
8.6.3
Set operational mode
The Rosemount 5408:SIS can be used as the level sensor in a BPCS or as a safety device in a
safety instrumented system.
If the Rosemount 5408:SIS is used as safety device in a Safety Instrumented System, then
the operational mode must be set to Safety (SIS). The Safety (SIS) operational mode can
be activated via the Guided Setup wizard, or as follows:
Use Rosemount Radar Master Plus
Procedure
1. Under
Configure
, select
Device Setup
→
Security
.
2. Under
Operational Mode
, select
Change
and follow the on-screen instructions.
Use AMS Device Manager and Field Communicator
Procedure
1. Select
Configure
→
Manual Setup
→
Device Setup
→
Security
.
2. Under
Safety Instrumented Systems
, select
Change Operational Mode
and follow the
on-screen instructions.
Note
When entering the Safety (SIS) operational mode, the analog output will be put into
alarm mode until the Safety Mode is enabled.
8.6.4
Enable safety mode
If the transmitter is configured for use in Safety (SIS), then the Safety Mode must be
enabled for the transmitter to become operational. When Safety Mode is enabled, the
transmitter is write protected (with or without a password) to prevent unauthorized
changes.
Use Rosemount Radar Master Plus
Procedure
1. Under
Configure
, select
Device Setup
→
Security
.
2. Under
Safety Mode
, select
Change
and follow the on-screen instructions.
Use AMS Device Manager and Field Communicator
Procedure
1. Select
Configure
→
Manual Setup
→
Device Setup
→
Security
.
2. Under
Safety Instrumented Systems
, select
Change Safety Mode
and follow the on-
screen instructions.
Reference Manual
Safety Instrumented Systems (SIS) requirements
00809-0100-4408
February 2020
Rosemount 5408 and 5408:SIS Level Transmitters
153
Содержание Rosemount 5408
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual February 2020 00809 0100 4408 viii Reference Manual ...
Страница 10: ...Introduction Reference Manual February 2020 00809 0100 4408 10 Reference Manual ...
Страница 20: ...Transmitter overview Reference Manual February 2020 00809 0100 4408 20 Reference Manual ...
Страница 72: ...Mechanical installation Reference Manual February 2020 00809 0100 4408 72 Reference Manual ...
Страница 100: ...Operation Reference Manual February 2020 00809 0100 4408 100 Reference Manual ...
Страница 240: ...Product certifications Reference Manual February 2020 00809 0100 4408 240 Reference Manual ...
Страница 242: ...Product certifications Reference Manual February 2020 00809 0100 4408 242 Reference Manual ...
Страница 244: ...Product certifications Reference Manual February 2020 00809 0100 4408 244 Reference Manual ...
Страница 246: ...Product certifications Reference Manual February 2020 00809 0100 4408 246 Reference Manual ...
Страница 248: ...Product certifications Reference Manual February 2020 00809 0100 4408 248 Reference Manual ...