![Emerson Rosemount 5408 Скачать руководство пользователя страница 143](http://html.mh-extra.com/html/emerson/rosemount-5408/rosemount-5408_reference-manual_103670143.webp)
3. Carefully pull the PTFE sealing straight out.
Reassembly procedures
Procedure
1. Clean the cavity with a lint-free cloth.
2. Verify the O-ring on the PTFE sealing is in place.
3. Gently insert the PTFE sealing until it stops, and then firmly push it all the way in.
Reference Manual
Service and troubleshooting
00809-0100-4408
February 2020
Rosemount 5408 and 5408:SIS Level Transmitters
143