Table 7-5: Status - Maintenance Required
(continued)
LCD display
message
Host diagnostic
message
Description
Recommended actions
HIGH
ALERT
High User Defined Alert
The user defined variable is above
the defined limit.
1. Bring the system to a safe
state.
2. Verify that the process variable
is within specified limits.
3. Reconfirm the user defined
alarm limit.
4. If not needed, disable this
alert.
LOW
ALERT
Low User Defined Alert
The user defined variable is below
the defined limit.
1. Bring the system to a safe
state.
2. Verify that the process variable
is within specified limits.
3. Reconfirm the user defined
alarm limit.
4. If not needed, disable this
alert.
VAR
OUTRNG
Linearized Variable Out
of Range
The level measurement is outside
the configured range for volume or
scaled variable, or both.
Accuracy of volume/scaled variable
measurement may be degraded.
1. If volume strapping table is
used, make sure level values
within operating range are
included.
2. If scaled variable table is used,
make sure input variable values
within operating range are
included.
DC
DEGRAD
Dielectric Constant
Estimation Degraded
The dielectric constant estimation is
degraded.
Accuracy of level measurement may
be degraded.
1. Check configuration of Bottom
Product Dielectric Constant.
2. Check configuration of
Reference Height and Bottom
Offset.
3. If not needed, disable Tank
Bottom Projection.
7.3
Troubleshooting guides
If there is a malfunction despite the absence of alerts, follow the procedures described in
the appropriate troubleshooting guide. Under each of the symptoms, specific suggestions
for solving problems are offered.
Related information
Incorrect level readings
Troubleshooting the 4-20 mA/HART Output
Reference Manual
Service and troubleshooting
00809-0100-4408
February 2020
Rosemount 5408 and 5408:SIS Level Transmitters
107
Содержание Rosemount 5408
Страница 4: ...4 ...
Страница 8: ...Contents Reference Manual February 2020 00809 0100 4408 viii Reference Manual ...
Страница 10: ...Introduction Reference Manual February 2020 00809 0100 4408 10 Reference Manual ...
Страница 20: ...Transmitter overview Reference Manual February 2020 00809 0100 4408 20 Reference Manual ...
Страница 72: ...Mechanical installation Reference Manual February 2020 00809 0100 4408 72 Reference Manual ...
Страница 100: ...Operation Reference Manual February 2020 00809 0100 4408 100 Reference Manual ...
Страница 240: ...Product certifications Reference Manual February 2020 00809 0100 4408 240 Reference Manual ...
Страница 242: ...Product certifications Reference Manual February 2020 00809 0100 4408 242 Reference Manual ...
Страница 244: ...Product certifications Reference Manual February 2020 00809 0100 4408 244 Reference Manual ...
Страница 246: ...Product certifications Reference Manual February 2020 00809 0100 4408 246 Reference Manual ...